Matching the Resolution of the Lens and Matrix of Medical Thermal Camera

The paper considers the interrelation between radius of the confusion lens circle and the period matrix of modulation transfer function (MPF) with the concurrence lens and the detector. We propose two criteria of the alignment lens and the detector. The first criterion is the equality of the lens and matrix on the level while the second criterion is the equality of the MPF lens detector at the Nyquist frequency. Finally, we uncover the reduced dependence of the normalized radius of the circle of confusion on the value of the contrast of the lens alignment

Publication year: 
2012
Issue: 
1
УДК: 
621.384.326
С. 117—120. Іл. 3. Бібліогр.: 6 назв.
References: 

1. Тарасов М.М., Якушенков Ю.Г. Инфракрасные системы “смотрящего типа”. — М.: Логос, 2004. — 444 с.
2. Хребтов И.А., Маляров В.Г. Неохлаждаемые тепловые матричные приемники ИК излучения // Оптический журнал. — 1997. — № 6. — С. 3—17.
3. Kaplan H. Practical applications of infrared and imaging equipment. — SPIE, Bellingham, Washington USA, 1999. — 164 p.
4. Иваницкий Г.В. Современное матричное тепловидение в биомедицине // Успехи физ. наук. — 2006. — 176, № 12. — С. 1295—1320.
5. Ллойд Дж. Системы тепловидения / Пер. с англ. — М.: Мир, 1978. — 416 с.
6. Колобродов В.Г., Лихоліт М.І. Проектування тепловізійних і телевізійних систем спостереження: Підручник. — К.: НТУУ “КПІ”, 2007. — 364 с.