modulation transfer function

Improvement of Calculating Medical Thermovision Camera Minimum Resolvable Temperature Difference

The paper is devoted to improvement of the equation for calculating the minimum resolvable temperature difference (MRTD) of medical thermovision camera by means of more sophisticated models of thermal images visual perception. An MRTD calculating algorithm, which is based on a more reliable approximation of the visual system modulation transfer function by Schultz, is considered. We obtain new expression for the bandwidth of Schultz’s approximation, which does not depend on the angular size of the Foucault pattern bar. An improved equation for calculating the MRTD was obtained.

Method for Assessing Modulation Transfer Function of Trifocal Intraocular Lens

The article investigates the quality of images formed by multifocal optical elements. The method for calculating of multifocal diffractive lens modulation transfer function (MTF) at presence of background is proposed. The method is based on a geometric calculation of the intensity of background images and the diffraction efficiency of the main image in each focal plane is considerated. The article has studied the contrast of main images formed by trifocal intraocular lens (IOL).