minimum resolvable temperature difference

Improvement of Calculating Medical Thermovision Camera Minimum Resolvable Temperature Difference

The paper is devoted to improvement of the equation for calculating the minimum resolvable temperature difference (MRTD) of medical thermovision camera by means of more sophisticated models of thermal images visual perception. An MRTD calculating algorithm, which is based on a more reliable approximation of the visual system modulation transfer function by Schultz, is considered. We obtain new expression for the bandwidth of Schultz’s approximation, which does not depend on the angular size of the Foucault pattern bar. An improved equation for calculating the MRTD was obtained.